Nearly half the score, and the part that separates this exam from a rhythm-reading course. The plan covers equipment care, machine settings, skin preparation, positioning, electrode placement for 12-lead, ambulatory, stress and telemetry recordings, special cases, checking that every lead recorded, fixing artifacts, mounting the tracing, and supporting the patient during a stress test.
Settings first. Standard speed is 25 mm per second and standard gain 10 mm per mV; a tracing at any other value has to say so, or it will be misread. Position the patient flat on the back, arms resting by the body and legs uncrossed; raise the backrest only for someone who cannot lie flat, and note it.
Chest leads, from landmarks you can feel:
- V1 goes in the fourth intercostal space at the right edge of the breastbone; V2 takes the same space at the left edge.
- V4 goes in the fifth intercostal space on the midclavicular line, and V3 halfway between V2 and V4.
- V5 and V6 continue level with V4, on the anterior and midaxillary lines.
To find the fourth space, start at the sternal angle, the ridge where the handle of the sternum meets its body. The second rib joins there; the space under it is the second, and you count down two more. On a patient with large breasts, place V4 to V6 beneath the breast tissue, not over it.
For a treadmill test, the limb electrodes move onto the torso, below the clavicles and on the lower abdomen or flanks. A Holter or event recorder uses its own sites, chosen over bone; telemetry uses three or five electrodes in set positions. Special cases include amputations, a pacemaker generator under the skin, children, and right-sided or posterior leads ordered by the provider. Whatever you change, note it on the tracing.
Reversed arm wires turn lead I upside down Swap the right and left arm electrodes and lead I comes out inverted, both P and QRS negative; aVR and aVL exchange tracings, and so do II and III. It can look like a heart on the wrong side. Check the limb clips before you send the tracing.
A stress test stops on symptoms, not on the schedule During exercise you watch the patient and the monitor together. Chest pain that grows, dizziness or near fainting, pale and clammy skin, a falling blood pressure, a sustained fast wide-complex rhythm, a monitor you can no longer read, or the patient asking to stop all mean you tell the supervising provider at once. The provider decides how the test ends; you do not push on to the next stage.